Model 910 Single beam photometer /
The Model OUM910 is microprocessor based and used with either a Model OUSAF10, AF11, DP11, or OUSAF12 online sensor to measure the spectral absorbance of process liquids in the visible and NIR region of the electromagnetic spectrum.
- Measures solids and absorbance up to 5 AU/OD units
- 4 X 20 character alphanumeric display and 6 button input for simple navigation
- Two 4-20mA analog outputs
- User-friendly menu based calibration: liquid free calibration using the EasyCal™ system and wet calibration using prepared liquid standards
- Four configurable digital inputs and three optically isolated digital outputs can be connected to a PLC for inclusion in a control scheme
- Built-in diagnostics to report any system errors or alarms, including sensor failures, open cable and internal processor exceptions
- Baseline and baseline shift functions allow operator to manually set artificial zero point
- Advanced log amp provides improved response characteristics
- Front panel environmental protection level is rated at NEMA 4/IP65 -optional NEMA 4/IP65 field mount enclosures available
Download |